CTD - Conductivity Temperature Depth : Micro CTD


CTD - Conductivity Temperature Depth : Micro CTD : Description


The Micro CTD is a small, high performance and flexible CTD. With rapid response times on conductivity (25 milliseconds at 1 m/s flow), temperature and pressure sensors, salinity spiking is minimized without relying on pumps or other moving parts. In this configuration, the Micro CTD comes with two additional sensors: the D&A OBS3 and the Seapoint Rhodamine Fluorometer. The CTD can be either self-powered or powered externally; similarly it can log data or communicate it in real-time to the surface via RS232. The unit comes with mounting brackets, a suspension bar, and interconnect cables between the additional sensors and the CTD. One single, consolidated data stream is provided.

CTD - Conductivity Temperature Depth : Micro CTD : Summary Information:


Name:Micro CTD
Category:CTD - Conductivity Temperature Depth
Model:Micro
Number:n/a
Description:The Micro CTD is a small, high performance and flexible CTD. With rapid response times on conductivity (25 milliseconds at 1 m/s flow), temperature and pressure sensors, salinity spiking is minimized without relying on pumps or other moving parts. In this configuration, the Micro CTD comes with two additional sensors: the D&A OBS3 and the Seapoint Rhodamine Fluorometer. The CTD can be either self-powered or powered externally; similarly it can log data or communicate it in real-time to the surface via RS232. The unit comes with mounting brackets, a suspension bar, and interconnect cables between the additional sensors and the CTD. One single, consolidated data stream is provided.
Configuration:Turbidity & Rhodamine
Application:For use in profiling the water column, particularly when high speed sampling is required.
Credibility Note:n/a
Case Study Link :n/a
Price:Consult factory


CTD - Conductivity Temperature Depth : Micro CTD : Usage Information:



Int Description:MC3 w/battery
Max Number Of Sensors:5
Depth Rating:500, 4500, 10000
Housing Material:Copolymer Acetal (Delrin), 316 Stainless Steel, Titanium
Weight Air:n/a
Weight Water:n/a
Length:243 (instrument, not including sensors) plus 272 (battery pack); can be placed end-to-end or side-by-side
Diameter:50 (instrument and battery pack)
Power Source :Battery
Housing Notes:Battery


CTD - Conductivity Temperature Depth : Micro CTD : Sensor Information:



Name:Conductivity
Manufacturer:Applied Microsystems
Type:Conductive
Model:n/a
Accuracy:0.01 mS/cm; 0.005 mS/cm precision
Range:0 - 7 S/m or 0 - 70 mS/cm
Response Time:25ms @ 1m/s
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution: 0.0015 mS/cm
Stabilityn/a
Location:n/a
Notes:The Applied Microsystems conductivity sensor is short, with a length of just under 3 cm. Sensor length plays a critical role in determining resistance to the introduction of new flows into the sensor cavity, and hence response times. The equation for response time is as follows:

Response Time = (0.85 * Length of Sensor ) / Flow Speed.

Using the above, our conductivity sensor has a response time of 25 milliseconds in a 1.0 m/second flow environment and 255 milliseconds in a 0.1 m/second flow environment. Given the above, no pumps or flow tubes are required.

Name:Pressure
Manufacturer:Keller
Type:Semiconductor bridge strain gauge
Model:PA-10
Accuracy:0.05%FS
Range:Various to 6000 dbar
Response Time:10ms
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution:0.002%FS
Stabilityn/a
Location:n/a
Notes:n/a

Name:Temperature
Manufacturer:Applied Microsystems
Type:Thermistor
Model:n/a
Accuracy:0.005C
Range:-2 to 32C
Response Time:100ms
Max. Frequency:25 Hz
Maximum Depth(m):n/a
Resolution:0.001C
Stabilityn/a
Location:n/a
Notes:Made of stainless steel

Name:Turbidity
Manufacturer:D&A Instruments
Type:Optical backscatter
Model:OBS3B or OBS3
Accuracy: 1.5 NTU's
Range: 2000 NTU's
Response Time:100 milliseconds
Max. Frequency:Varies by CTD model (25Hz Micro, 10Hz Smart, 8Hz Plus or EMP)
Maximum Depth(m):500 / 2000
Resolution:0.1 NTU's
Stabilityn/a
Location:n/a
Notes:500m version available in Delrin housing; 2000m version available in anodized aluminum

Name:Rhodamine
Manufacturer:Seapoint Sensors
Type:Flurometric (LED)
Model:SRF (No Pump)
Accuracy:V/Ug/l: 1.0, 0.33, 0.1, 0.033 (see associated Ranges below; see associated Gain in Notes)
Range:ug/l: 5, 15, 50, 150 (see associates sensitivities in Accuracy above; see associated Gain in Notes below)
Response Time:0.1 s
Max. Frequency:Varies by CTD model (25Hz Micro, 10Hz Smart, 8Hz Plus)
Maximum Depth(m):6000m
Resolution:n/a
Stabilityn/a
Location:With any CTD model
Notes:Low power consumption for in-situ applications; Gains of 30x, 10x, 3x, 1x apply to above accuracies and ranges