CTD - Conductivity Temperature Depth : EMP 2000


CTD - Conductivity Temperature Depth : EMP 2000 : Description


The EMP 2000 is a field proven instrument capable of monitoring a wide variety of water quality parameters. The EMP-2000 comes standard with conductivity and pressure sensors, and can support as many as 10 additional sensors (see Sensor Information for the list of available sensors). The instrument is capable of deployments to 300m.

CTD - Conductivity Temperature Depth : EMP 2000 : Summary Information:


Name:EMP 2000
Category:CTD - Conductivity Temperature Depth
Model:n/a
Number:n/a
Description:The EMP 2000 is a field proven instrument capable of monitoring a wide variety of water quality parameters. The EMP-2000 comes standard with conductivity and pressure sensors, and can support as many as 10 additional sensors (see Sensor Information for the list of available sensors). The instrument is capable of deployments to 300m.
Configuration:-
Application:Ideal for environmental impact studies, aquaculture, watershed monitoring or harbour monitoring, the EMP 2000 can be used in-situ or for coastal profiles.
Credibility Note:n/a
Case Study Link :n/a
Price:Consult factory


CTD - Conductivity Temperature Depth : EMP 2000 : Usage Information:



Int Description:ECM
Max Number Of Sensors:8
Depth Rating:2000
Housing Material:6061-T6 Aluminum Hard Anodized
Weight Air:23.6
Weight Water:8.9
Length:1180 (overall length including sensors & shackle)
Diameter:178
Power Source :Internal
Housing Notes:n/a


CTD - Conductivity Temperature Depth : EMP 2000 : Sensor Information:



Name:Conductivity
Manufacturer:Applied Microsystems
Type:Conductive
Model:n/a
Accuracy:0.01 mS/cm; 0.005 mS/cm precision
Range:0 - 7 S/m or 0 - 70 mS/cm
Response Time:25ms @ 1m/s
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution:0.0015 mS/cm
Stabilityn/a
Location:n/a
Notes:The Applied Microsystems conductivity sensor is short, with a length of just under 3 cm. Sensor length plays a critical role in determining resistance to the introduction of new flows into the sensor cavity, and hence response times. The equation for response time is as follows:

Response Time = (0.85 * Length of Sensor ) / Flow Speed.

Using the above, our conductivity sensor has a response time of 25 milliseconds in a 1.0 m/second flow environment and 255 milliseconds in a 0.1 m/second flow environment. Given the above, no pumps or flow tubes are required.

Name:Dissolved Oxygen
Manufacturer:YSI
Type:Polarographic
Model:n/a
Accuracy: 0.2 mg/l
Range:0 to 15 mg/l
Response Time:60s
Max. Frequency:Varies by CTD model (25Hz Micro, 10Hz Smart, 8Hz Plus)
Maximum Depth(m):300
Resolution:0.01 mg/l
Stabilityn/a
Location:n/a
Notes:n/a

Name:pH
Manufacturer:Innovative Sensors
Type:Combination Electrode
Model:A23B-I1B2Z-Z1HZ
Accuracy: 0.1 pH
Range:2 to 12 pH
Response Time:15 seconds
Max. Frequency:Varies by CTD model (25Hz Micro, 10Hz Smart, 8Hz Plus or EMP)
Maximum Depth(m):2000
Resolution:0.01 pH
Stabilityn/a
Location:n/a
Notes:n/a

Name:ORP / Redox
Manufacturer:Innovative Sensors
Type:Combination Electrode
Model:A23B-E1P0Z-D1HZ
Accuracy:± 2 mV
Range:± 2000 mV
Response Time:15 seconds
Max. Frequency:25-10-8
Maximum Depth(m):n/a
Resolution:1 mV
Stabilityn/a
Location:n/a
Notes:Requires CTD or EMP

Name:Current
Manufacturer:Marsh McBirney
Type:Two Axis Electromagnetic 1.5"
Model:n/a
Accuracy:±2% of reading
Range: ± 300cm/s
Response Time:5 s
Max. Frequency:Varies by instrument model (10Hz Smart or 8Hz on TWG and EMP)
Maximum Depth(m):n/a
Resolution:0.1 cm/s
Stabilityn/a
Location:n/a
Notes:n/a

Name:Pressure
Manufacturer:Keller
Type:Semiconductor bridge strain gauge
Model:PA-10
Accuracy:0.05%FS
Range:Various to 6000 dbar
Response Time:10ms
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution:0.005%FS
Stabilityn/a
Location:n/a
Notes:n/a

Name:Turbidity
Manufacturer:D&A Instruments
Type:Optical backscatter
Model:OBS3B or OBS3
Accuracy: 1.5 NTU's
Range: 2000 NTU's
Response Time:100 milliseconds
Max. Frequency:Varies by CTD model (25Hz Micro, 10Hz Smart, 8Hz Plus or EMP)
Maximum Depth(m):500 / 2000
Resolution:0.1 NTU's
Stabilityn/a
Location:n/a
Notes:500m version available in Delrin housing; 2000m version available in anodized aluminum

Name:Heading
Manufacturer:KVH
Type:Flux Gate
Model:KVHC100
Accuracy:+/-0.5 degrees
Range:0 to 360 degrees
Response Time:100 milliseconds
Max. Frequency:Varies by instrument model (10Hz on Smart & ECM; 8Hz on TWG or EMP)
Maximum Depth(m):6000m
Resolution:0.1 degrees
Stabilityn/a
Location:n/a
Notes:n/a

Name:PAR
Manufacturer:Licor
Type:Photo diode
Model:LI-192SA
Accuracy:± 5% in air traceable to National Bureau of Standards
Range:0-3055 umols/sm2 (micromole)
Response Time:10 microseconds
Max. Frequency:n/a
Maximum Depth(m):340m
Resolution:1 umol/sm2 (micromole)
Stabilityn/a
Location:n/a
Notes:Plus; Smart; Micro


Click here for the full list of Applied Microsystems products

 

 

Applied Microsystems Ltd. 2071 Malaview Ave.W. Sidney, BC, Canada V8L 5X6
Contact Us: sales@AppliedMicrosystems.com Tel:+1-250-656-0771 Fax:+1-250-655-3655