CTD - Conductivity Temperature Depth : Micro CTD


CTD - Conductivity Temperature Depth : Micro CTD : Description


The Micro CTD is a small, high performance and flexible CTD. With 25Hz sampling and conductivity response times of 25 milliseconds at 1 m/s flow, the instrument is ideal for integration onto high speed vehicles. With rapid response times on all three sensors, salinity spiking is minimized without relying on pumps or other moving parts. This configuration of Micro CTD is externally powered; other self-powered configurations are available.

CTD - Conductivity Temperature Depth : Micro CTD : Summary Information:


Name:Micro CTD
Category:CTD - Conductivity Temperature Depth
Model:Micro
Number:n/a
Description:The Micro CTD is a small, high performance and flexible CTD. With 25Hz sampling and conductivity response times of 25 milliseconds at 1 m/s flow, the instrument is ideal for integration onto high speed vehicles. With rapid response times on all three sensors, salinity spiking is minimized without relying on pumps or other moving parts. This configuration of Micro CTD is externally powered; other self-powered configurations are available.
Configuration:-
Application:Ideal for integration onto AUVs, ROVs, or tow bodies, particularly in high-speed applications.
Credibility Note:Integrated on Brooke Ocean Technologies MVP (Moving Vessel Profiler) and multiple AUVs / ROVs
Case Study Link :n/a
Price:Contact factory


CTD - Conductivity Temperature Depth : Micro CTD : Usage Information:



Int Description:MC3
Max Number Of Sensors:5
Depth Rating:500, 4500, 10000
Housing Material:Copolymer Acetal (Delrin), 316 Stainless Steel, Titanium
Weight Air:n/a
Weight Water:n/a
Length:243 (not including sensors)
Diameter:50
Power Source :External
Housing Notes:Micro CTD is modular and can function with or without the battery pack. Multiple battery packs can be daisy chained together to provide additional power capacity.


CTD - Conductivity Temperature Depth : Micro CTD : Sensor Information:



Name:Conductivity
Manufacturer:Applied Microsystems
Type:Conductive
Model:n/a
Accuracy:0.01 mS/cm; 0.005 mS/cm precision
Range:0 - 7 S/m or 0 - 70 mS/cm
Response Time:25ms @ 1m/s
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution: 0.0015 mS/cm
Stabilityn/a
Location:n/a
Notes:The Applied Microsystems conductivity sensor is short, with a length of just under 3 cm. Sensor length plays a critical role in determining resistance to the introduction of new flows into the sensor cavity, and hence response times. The equation for response time is as follows:

Response Time = (0.85 * Length of Sensor ) / Flow Speed.

Using the above, our conductivity sensor has a response time of 25 milliseconds in a 1.0 m/second flow environment and 255 milliseconds in a 0.1 m/second flow environment. Given the above, no pumps or flow tubes are required.

Name:Pressure
Manufacturer:Keller
Type:Semiconductor bridge strain gauge
Model:PA-10
Accuracy:0.05%FS
Range:Various to 6000 dbar
Response Time:10ms
Max. Frequency:25Hz
Maximum Depth(m):6000m
Resolution:0.002%FS
Stabilityn/a
Location:n/a
Notes:n/a

Name:Temperature
Manufacturer:Applied Microsystems
Type:Thermistor
Model:n/a
Accuracy:0.005C
Range:-2 to 32C
Response Time:100ms
Max. Frequency:25 Hz
Maximum Depth(m):n/a
Resolution:0.001C
Stabilityn/a
Location:n/a
Notes:Made of stainless steel


Click here for the full list of Applied Microsystems products

 

 

Applied Microsystems Ltd. 2071 Malaview Ave.W. Sidney, BC, Canada V8L 5X6
Contact Us: sales@AppliedMicrosystems.com Tel:+1-250-656-0771 Fax:+1-250-655-3655